ROX
OPTICALDISSOLVEDOXYGENCALIBRATIONTIPS
ROX FACTS
The ROX Optical Dissolved Oxygen sensor does not require any special sonde setup or burn-in.
Calibration data is stored in the ROX so it can be calibrated in one sonde and then used in another
without recalibrating the probe in the new sonde.
Calibration data is automatically transferred to the host sonde as soon as the sonde powers up the
ROX sensor.
The ROX has excellent long term stability and will achieve ± 1% accuracy when calibrated in a
(temperature-stable) saturated water or saturated air environment.
Field DO calibrations should be avoided!
Remove the wiper when the sensor is going to be exposed to sodium sulfite (zero DO test) or
Formazin-based turbidity standards.
The ROX sensor must remain hydrated at all times! See Appendix M in the 6-Series manual for
additional information.
CALIBRATION METHOD – ONE-POINT “SATURATED AIR”
1. Dry the temperature sensor and the ROX membrane to remove any water droplets.
2. Put 1 inch of water in the calibration cup and thread onto the endcap ½ thread.
Do not fully thread the calibration cup onto the endcap!
3. Allow the sonde to sit in the calibration cup for 15 minutes to ensure that the air is 100%
saturated and that the temperature is stable.
4. Select ODO% and enter the barometric pressure in mmHg (inches x 25.4) and press Enter. The
sonde will do a 4 second count down and then say “Press Enter when the readings are stable”.
Wait at least 30 seconds before you press the Enter button.
Note: the sonde will only read 100% saturation if the barometric pressure is 760 mmHg. To
determine the actual DO in %, divide the barometer reading by 760 and then multiply that
number by 100.
Example: 750 ÷ 760 = 0.9868 x100 = 98.68%
Check the ODO Gain in the Advanced Menu. An acceptable range is 0.85 to 1.15, NOTE:
Maximum range before getting error message is 1 ± .25.
CALIBRATION METHOD – ONE-POINT “SATURATED WATER”
1. Using a medium-size aquarium air pump and air stone, sparge a five-gallon bucket of tap water
for 1 hour.
2. Place the assembled sonde in the bucket and allow the sonde to equilibrate to the water
temperature for 5 minutes.
3. Select ODO% and enter the barometric pressure in mmHg (inches x 25.4) and press Enter. The
sonde will do a four-second countdown and then say “Press Enter when the readings are
stable”. Wait at least 30 seconds before you press the Enter button.
(continued)