3. At the recovery, the process is repeated. The sonde, when recovered, would be externally wiped
clean (not the sensors) and then wrapped again in the wet towel and placed back in the white cooler.
The DO data now collected by the sonde will be used as part of the post calibration record. By
having this information recorded on the actual data set the QA will always be available for future
reference.
4. The other sensors will be post-calibrated when the sonde is returned to the lab. These tests will
determine what effect (if any) fouling might have had on the data. When the post-calibration is
completed the sonde can be returned to the saturated water bath for a final verification.