SPECIFICATIONS
Display Range Accuracy Resolution
Conductivity,
Auto-ranging
0.0 to 499.9 uS/cm
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm
±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm
±1% of reading plus 0.05 uS/cm
±2.5% of reading plus 0.5 mS/cm
0.01 uS/cm
1 mS/cm
0.01 mS/cm
0.1 mS/cm
Salinity 0.0 to 70.0 ppt 0.2% Full Scale 0.1 ppt
Temperature
°C
-10.0 to 90 °C ±0.2 °C or ±0.4% Full Scale,
whichever is greater
0.1 °C
Reference Temperature
15.0 to 25.0
°C
Temperature Coefficient 0.0% to 4.0%
Cell Constant
5.00
± 0.50
TDS Constant Range 0.30 to 1.00
Power/Battery life
One 9V batter
y
Approximately 500 hour
Calibration Back-up Yes
Audio Feedback Yes, on all touch keys
Instrument Case
Waterproof when connector cap installed, IP 67
Operating Temp. Range
0 to 50
°C
Operating Relative Humidity Range up to 95%
Temperature Probe
Thermistor, 10k / 25
°C
Dimensions (L x W x H)
18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)
Weight (batteries included)
270 grams (.6 lb)
RECOMMENDED SPARE PARTS LIST
PART # DESCRIPTION
300-1 1-meter probe and cable assembly
300-4 4-meter probe and cable assembly.
300-10 10-meter probe and cable assembly.
606043 Carrying case, hard sided.
485 Carrying case, soft sided.
Item #606042REF
Revision A, July 2012
For the latest version of this manual, visit www.ysi.com