YSI Model 55 Handheld Dissolved Oxygen and Temperature System Operations Manual
64.4. Membrane Installation A B E F C D A. Secure a membrane between your thumb and the probe body. Add electrolyte to the probe until
74.5. Probe Operation and Precautions 1. Membrane life depends on usage. Membranes will last a long time if installed properly and treated with ca
85. Calibration Dissolved oxygen calibration must be done in an environment with a known oxygen content. Since the amount of oxygen in the atmosphe
95. When the proper altitude appears on the LCD, press the ENTER key. The Model 55 should now display CAL in the lower left of the display, the cal
106. Principles Of Operation The sensor consists of an acrylic body with a circular gold cathode embedded in the end. Inside the gold ring there is
11Type 3 Errors A. Altitude: Operator Error: A 1000 foot error in altitude (when calibrating) is equal to an error of approximately 3.6% at the 1
127. Troubleshooting NOTE: An error displayed briefly during the first few seconds after turning the instrument on does NOT indicate a problem. S
13 SYMPTOM POSSIBLE CAUSE ACTION 9. Main display reads "Er 1" or Main display reads “Err” (Secondary display reads “ra”) A. Instrument
148. Warranty And Repair YSI Model 55 Dissolved Oxygen and Temperature Meters are warranted for two years from date of purchase by the end user aga
158.1. Cleaning Instructions NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned an
168.2. Packing Instructions 1. Clean and decontaminate items to insure the safety of the handler. 2. Complete and include the Cleaning Certificat
179. Required Notice The Federal Communications Commission defines this product as a computing device and requires the following notice: This equip
1810. Accessories And Replacement Parts The following parts and accessories are available from YSI or any Franchise Dealer authorized by YSI. YSI
1911. Appendix A - Solubility Table Solubility of Oxygen in mg/L in Water Exposed to Water-Saturated Air at 760 mm Hg Pressure. Salinity = Measure
20 Temp oC Chlorinity: 0 Salinity: 0 5.0 ppt 9.0 ppt 10.0 ppt 18.1 ppt 15.0 ppt 27.1 ppt 20.0 ppt 36.1 ppt 25.0 ppt 45.2 ppt 20.0 9.09
2112. Appendix B - Conversion Chart To Convert From To Equation Feet Meters Multiply by 0.3048 Meters Feet Multiply by 3.2808399 Degrees
1725 Brannum Lane Yellow Springs, Ohio 45387 USA 055207 937 767-7241 • 800 765-4974 • Fax 937 767-9353 A55207D - Web [email protected] •
iTable of Contents 1. General Description ...1 2.
11. General Description The YSI Model 55 Handheld Dissolved Oxygen System is a rugged, micro-processor based, digital meter with an attached YSI di
22. Specifications Probe Operating Environment Medium: fresh, sea, or polluted water Temperature: -5 to +45oC Depth: 0 to 12, 0 to 25 or 0 t
33. Preparing The Meter 3.1. Unpacking When you unpack your new YSI Model 55 Handheld Dissolved Oxygen System for the first time, check the packin
4Turn the instrument on by pressing and releasing the ON/OFF button on the front of the instrument. The instrument will activate all segments of the
54. Preparing The Probe 4.1. Description The YSI Model 55 dissolved oxygen probe is a non-detachable, polarographic sensor designed specifically f